H P CODEMASTER

H P CodeMaster™ XL  The technical data given in this publication is for general information and are subject to change without notice. Actual configuration on the unit may vary. Contact our sales representatives for a complete list of details.

Specifications Dimensions 7.9”H x 11.8”W x 15.6”L/20 x 30 x 39.7 cm)

 Weight 24 lbs./10.9 kg

Battery Type: Rechargeable, sealed lead-acid battery. 4 Ah, 12 V nominal.

Charge time: 2 hours to 90% of full capacity. 18 hours to 100% capacity.

 Repeated charging to less than 100% will reduce the useful life of the battery.

 Capacity: 2.5 hours monitoring or 50 full-energy discharges or 1 hour monitoring and recording.

 AC input: 100 to 230 VAC ±15%, 50/60 Hz. Battery

Defibrillator

 Waveform: damped sinusoidal.

Output energy (delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and 360 joules.

Charge control: push-button on apex paddle and on front panel.

 Charge time: less than 5 seconds to 360 joules with a functional battery installed.

 Charge-done indicator: display indicates available energy, charge-done light on apex paddle, charge done tone.

Synchronizer: SYNC message appears on the monitor and is annoted periodically on the recorder while in synchronous mode.

An audible beep sounds with each detected R-wave, while a marker on the monitor and sync mark on the recorder strip indicate the discharge point.

 Paddles: standard paddles are anterior/anterior.

 Adult electrodes (83 cm sq) slide off to expose pediatric electrodes (21 cm sq).

Paddle cord is 10 feet (3m).

 Full range of internal paddles available

 Monitor

Input: ECG activity may be viewed through the paddles or patient cable.

 Lead fault: LEADS OFF message and dashed baseline appear on the monitor if a lead becomes disconnected.

 Common mode rejection: greater than 100 dB measured as per AAMI standards for cardiac monitors.

 Heart rate display: digital read out on monitor from 20 to 280 bpm.

Recorder

 Annotations: time, date, heart rate, event markers, ECG mode, and selected energy.

 Recorder mode: automatically documents events and ECG during defibrillation episodes.

The recorder can be configured to run in either real time or with a 6-second delay.

Frequency response: 0.5 to 40 Hz monitor bandwidth. 0.05 to 150 Hz diagnostic bandwidth is selectable